Operating Conditions (ColorFlex)

ColorFlex can be stored in an area with a temperature range of -10°F to 140°F (-23°C to 60°C) and can be operated under temperature conditions of 50°F to 110°F (10°C to 43°C).  The ColorFlex may be operated in humidity conditions from 5% to 85%.  Do not leave ColorFlex in an area where temperature or humidity extremes are possible.

 

Fluctuating temperatures may produce unstable readings, as the silicon detector is sensitive to rapid temperature changes.  If the humidity is consistently above 80% for an extended period, long-term performance may be affected by oxidation (rusting) of electrical contacts.  At humidities below 20%, static electricity may build up, so care should be taken that users are well grounded before using the instrument, or the instrument should be placed on a static mat.

 

 

If additional assistance is required, you may contact HunterLab.