Do HunterLab sphere instruments conform to ASTM D1003?
Yes, but note there are two different approved instrument geometries cited in the ASTM 1003 method.
How can I upgrade or install EasyMatch QC software on a new computer?
For either a network or local PC installation, the person installing EasyMatch QC software must be a System Administrator for the PC, with software installation privileges for both the old and new computers.
How can I upgrade EasyMatch QC software on an existing computer?
For either a network or local PC installation, the person installing EasyMatch QC software must be a System Administrator for the PC, with software installation privileges for that computer.
Is there an Internet Site for converting and displaying color values?
FAQ: “… Was wondering – do you know of a web site or application where if CIELAB values were entered, the corresponding color would be displayed? Can I also convert from one color scale to another?”
What Fuse Protection is in place with HunterLab Instruments?
All HunterLab instruments have fuses built into the main circuitry to protect the instrument against unusual and unexpected changes in voltage that could damage the electrical components.
Can Power Supplies in HunterLab Instruments be used everywhere?
FAQ: “We need to take our LabScan XE instrument to Germany for some plant trial testing. Our machine is a US spectrophotometer 110v unit and, obviously, the power supply in Germany is 220v. The label on the back of the instrument says ‘110v-220v’, which sounds a little ambiguous to me. Do we need a voltage […]
Optimal inter-instrument agreement and communication of color values
To communicate color values effectively, completely specify the 6 parameters of your measurement method and verify that your color measurement is repeatable. These are the key factors to ensure inter-instrument agreement. The best inter-instrument agreement in color measurement at multiple locations occurs when a product is measured with the same instrument model from the same […]
How do I measure Haze with EasyMatch QC?
Haze is the scattering of light by a nearly clear specimen that results in a slightly cloudy appearance when the sample is viewed in transmission.
How do I perform an DQ IQ OQ PQ RQ using EasyMatch QC-ER?
What is the Hunter Rd, a, b Color Scale?
The Hunter Rdab opponent-color scale, developed in 1943, was an early version of the Hunter L, a, b scale. Instead of L, Rd is used to designate lightness and is identical to the CIE Y Brightness/Luminosity value. Sometimes the Rd, a, b color scale is shown as LRd, aRd, bRd.
If I select D65/10 to read XYZ what will happen when I read YI?
FAQ: “One question we are a little unclear about – if you set X, Y, Z on D65/10° in the Color Data View to read back the White Tile for a calibration check, will it make the YI Yellowness Index read in D65/10°, or is it fixed to C/2°?”